Tuan M Vo, Toan T Nguyen, Dat Q Nguyen, T. N. Tran, Yunhee Kang
2026.2.5Expert Review of Medical Devices
tlooto Summary
The Fitbit Charge 6 did not show reliable correlation with the subjective sleep data reported by one-month postpartum breastfeeding women, suggesting the validity and reliability of sleep data collection methods for this demographic is still questionable.
Abstract
BACKGROUND A new health monitoring device called Fitbit Charge 6 has appeared on the Vietnamese market. This study aimed to evaluate the consistency between Fitbit Charge 6 and structured interview questions in measuring sleep quality among breastfeeding mothers.
RESEARCH DESIGN AND METHODS A cross-sectional study utilizing random convenience sampling was conducted in Ho Chi Minh City, Vietnam, involving 208 breastfeeding women one month postpartum. The participants used the Fitbit Charge 6 device continuously for four days and three nights to gather daily sleep information; then, they were administered a structured sleep questionnaire via home interviews.
RESULTS The correlation between the total sleep time (TST) recorded by both the device and interview was weak (r = 0.28), as was the correlation for short nap duration (r = 0.25). The agreement between the Fitbit Sleep Score and sleep satisfaction from the questionnaire was 79.9%, with a notably low kappa coefficient (0.02), indicating a lack of strong agreement.
CONCLUSION The Fitbit Charge 6 did not show reliable correlation with the subjective sleep data reported by one-month postpartum breastfeeding women. Further research employing rigorous comparative studies is necessary to assess the validity and reliability of sleep data collection methods for this demographic.
Citation format
VO, Tuan M, et al. Investigation of consistency in sleep quality evaluation using a fitbit device compared to a structured questionnaire among breastfeeding mothers in ho chi minh city, vietnam. Expert Review of Medical Devices, 2026, 23(4): 439–447.