Wenyan Zhang, Zelong Chen, Q. Tao, Caixia Li, Jianping Wang, Yizhe Zhang, D. Xie, Jianhua Ma, T. Niendorf, Yanqiu Feng
2026.2.27MAGNETIC RESONANCE IN MEDICINE
tlooto Summary
This study demonstrates the feasibility of 3D whole kidney T1 mapping using dual breath-hold Look-Locker T1-weighting in conjunction with balanced SSFP readouts and dictionary matching, and provides a technical foundation for improving the understanding of renal (patho-)physiology.
Abstract
PURPOSE To develop and validate dual breath-hold Look-Locker inversion recovery (Dual 3DLL) MRI for 3D whole-kidney T1 quantification using dictionary matching. METHODS Dual 3DLL employs two breath-hold Look-Locker prepared acquisitions using nonselective adiabatic inversion in conjunction with 3D balanced SSFP readouts at three inversion times for each breath-hold. Retrospective registration of T1-weighted images was implemented to reduce misalignment between two breath-holds. For T1 quantification, Bloch equation-based dictionary matching was applied. Dual 3DLL was validated in phantoms using inversion recovery spin-echo (IRSE) based T1 mapping as a reference and in 11 healthy subjects, where 2D MOLLI-based renal T1 mapping was used as a reference. RESULTS Validation in the phantom showed an agreement between Dual 3DLL and IRSE (precision: coefficient of variation = 2.5%, relative error compared to IRSE: 4.3% ± 2.0%). The human feasibility study demonstrated the clinical applicability of Dual 3DLL. Averaged T1 derived from whole-kidney coverage Dual 3DLL was in accordance with that of 2D MOLLI (renal cortex: T1,Dual 3DLL = 1413.41 ± 114.73 ms, T1,MOLLI = 1394.39 ± 58.01 ms; medulla: T1,Dual 3DLL = 1827.48 ± 108 ms, T1,MOLLI = 1843.67 ± 91.54 ms). The coefficients of variation between T1 obtained from Dual 3DLL and MOLLI were 6.1% for the renal cortex and 4.4% for the medulla. CONCLUSION This study demonstrates the feasibility of 3D whole kidney T1 mapping using dual breath-hold Look-Locker T1-weighting in conjunction with balanced SSFP readouts and dictionary matching, and provides a technical foundation for improving our understanding of renal (patho-)physiology.
Citation format
ZHANG, Wenyan, et al. 3D whole-kidney t1 mapping using look-locker inversion recovery in conjunction with balanced SSFP readout and dictionary matching. MAGNETIC RESONANCE IN MEDICINE, 2026, 96(1): 109–120.