C. Thorne, A. Gatt, C. DeRaffaele, Geoffrey Attard, N. Papanas, C. Formosa
tlooto Summary
The new device provides pressure measurements comparable to the FScan system and appears promising for monitoring foot health.
Abstract
This study aims to evaluate the performance of an innovative in-shoe sensing device against the FScan system. Across five trials, the novel device demonstrated consistently strong correlations with FScan peak pressure readings (r = 0.801, r = 0.978, r = 0.813, r = 0.887, r = 0.944). Superimposed peak-pressure plots revealed highly similar waveform patterns between the two systems. Error-based metrics also supported this equivalence. Root Mean Squared Logarithmic Error values similarly suggested consistent predictive agreement. Temperature-sensing performance was assessed by comparing in-shoe sensor readings by the novel device with thermal-camera measurements. The thermal camera detected an average temperature change of 3.7°C, whereas the in-shoe sensor recorded an average change of 0.67°C. There was substantial variability in both systems, with higher variation observed in the in-shoe sensor than in the thermal camera. Despite variability, temperature and pressure measurements from the novel device were strongly correlated (r = 0.87). In conclusion, the new device provides pressure measurements comparable to the FScan system and appears promising for monitoring foot health.
Citation format
THORNE, C., et al. Development and validation of a wireless, low-cost device for dual measurement of in-shoe plantar pressure and temperature in high-risk diabetic feet. International Journal of Lower Extremity Wounds, 2026: 15347346261428561.