MedicineEngineering

D. J. English, N. Weerakkody, A. Zacharias, R. A. Green, M. de Noronha, C. Hocking, A. Kumar, X. Li, R. Rico Bini

2026.1.29International Biomechanics

DOI: 10.1080/23335432.2026.2624302

tlooto Summary

The head-mounted laser is appropriate for clinical movement assessment in the sagittal and transverse planes but not recommended for the coronal plane, and other devices should be considered for complete clinical assessment of cervical position sense.

Abstract

ABSTRACT The head-mounted laser is commonly used in clinical proprioceptive tests, but its ability to measure movement across three planes of motion has not been investigated. Therefore, this study evaluated the head-mounted laser against a validated gold standard multi-sensor inertial measurement unit (IMU) (XsensTM) in measuring cervicocephalic movement across three planes. Fourteen healthy adults (seven males and seven females) performed six repetitions each of active cervical flexion, extension, lateral flexion and rotation, whilst instructed to maintain the laser beam within a bullseye corresponding to 4.5° of movement. Primary plane means XsensTM range of motion (ROM) was used to evaluate if movement remained less than 4.5°. During flexion, extension and rotation movements, mean ROM in the primary plane using XsensTM remained within the 4.5° threshold (2.97°−3.57°), indicating that the head-mounted laser corresponded with the IMU (p ≤ 0.01). However, XsensTM mean lateral flexion movement reached 11.34° and 12.10° for left and right lateral flexion, indicating poor correspondence (p = 0.95 and 0.97). The head-mounted laser is appropriate for clinical movement assessment in the sagittal and transverse planes but not recommended for the coronal plane. Other devices should be considered for complete clinical assessment of cervical position sense.

Citation format

ENGLISH, D. J., et al. Is the head-mounted laser an appropriate tool to measure cervical movement across three planes? International Biomechanics, 2026, 13(1): 1–11.