2019.1.1International Journal of Exercise Science

DOI: 10.70252/abej8080

tlooto Summary

If visual jump determination of SBP cannot be avoided, re-assessment using a more traditional method (i.e. auscultation) is recommended.

Abstract

Purpose This study was designed to investigate differences in systolic blood pressure measurements as obtained through auscultation and observation of the visual jump on the manometer. Methods Men (n = 21; 26.9 ± 7.4 yrs) and women (n = 22; 29.3 ± 13.9 yrs) volunteered to have resting systolic blood pressure (SBP) assessments. During the same cardiac inflation-deflation cycle of traditional sphygmomanometry, the initial visual jump of the manometer needle and first Korotkoff sound heard were recorded. Duplicate assessments were made in each arm with 30 sec between intra-arm trials. Results Paired t-test results indicated there were no within-method differences between arms for visual jump (R: 132.1 ± 11.3; L: 131.8 ± 10.5 mmHg) or auscultation (R: 116.8 ± 9.0; L: 113.5 ± 8.8 mmHg). There were methodological differences within arm with visual jump being the higher of the two (right: t(42) = −12.69; left: t(42) = −11.37; p < .001). Conclusion If visual jump determination of SBP cannot be avoided, re-assessment using a more traditional method (i.e. auscultation) is recommended.

Citation format

TRIGG, Sam, et al. Comparison of systolic blood pressure measurements by auscultation and visual manometer needle jump. International Journal of Exercise Science, 2019, 12: 214–220.