Jie Lin, Yongsheng Liu, Yueheng Yang, Zhaochu Hu
2016.6.1Solid Earth Sciences
Abstract
Abstract LA-ICP-MS and LA-MC-ICP-MS have been the techniques of choice for achieving accurate and precise element content and isotopic ratio, the state-of-the-art technique combines the advantages of low detection limits with high spatial resolution, however, the analysis accuracy and precision are restricted by many factors, such as sensitivity drift, elemental/isotopic fractionation, matrix effects, interferences and the lack of sufficiently matrix-matched reference materials. Thus, rigorous and suitable calibration and correction methods are needed to obtain quantitative data. This review systematically summarized and evaluated the interference correction, quantitative calculation and sensitivity correction strategies in order to provide the analysts with suitable calibration and correction strategies according to the sample types and the analyzed elements. The functions and features of data reduction software ICPMSDataCal were also outlined, which can provide real-time and on-line data reduction of element content and isotopic ratios analyzed by LA-ICP-MS and LA-MC-ICP-MS.
Citation format
LIN, Jie, et al. Calibration and correction of LA-ICP-MS and LA-MC-ICP-MS analyses for element contents and isotopic ratios. Solid Earth Sciences, 2016, 1: 5–27.