Tang Da-hai, Wu Ling, He Jia-lei
2008.6.28Chinese Journal of Nautical Medicine and Hyperbaric Medicine
tlooto Summary
The SPOTCHEM-4410 analysator is accurate, stable and has strong resistance against interference in medical transport ship environment.
Abstract
Objective To evaluate the clinical application analysis performance of the SPOTCHEM-4410 analysator in medical transport ship environment. Methods Detection results in hospital laboratory and medical ship during navigation were measured,then accuracy and precision of the SPOTCHEM-4410 analysator were evaluated, respectively. Results There were positive correlation of total protein (TP), albumin (Alb), bilirubin total (T-Bil), glutamate-pyruvate transaminase (GPT), glutamate-oxalacetic transaminase (GOT) and lactate dehydrogenase( LDH) in the tests before and after navigation ( r > 0. 975, P > 0. 05 ). The error of measurement in medical ship during navigation were lower than 1/4 orl/3 of the CLJA' 88 permissible errors, which was permitted by precision requirement. Conclusions The SPOTCHEM-4410 analysator is accurate, stable and has strong resistance against interference. Key words: Dry reagent chemistry analysator; Medical transport ship; Clinical analysis application
Citation format
DA-HAI, Tang; LING, Wu; JIA-LEI, He. Application of the SPOTCHEM-4410 analysator in medical transport ship environment. Chinese Journal of Nautical Medicine and Hyperbaric Medicine, 2008, 15: 172–173.