Open AccessPhysicsMedicine

Joseph Busher, Edith Touchet-Valle, Chenhao Sun, S. Wright, M. McDougall

2023.4.10CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING

DOI: 10.1155/2023/7137889

Abstract

The utility of interleaved odd-number leg birdcage coils is demonstrated for decoupling in double- and triple-tuned multinuclear applications. The birdcage was designed to geometrically decouple from a planar double-tuned (1H-23Na) array and from a 31P saddle coil insert to create a triple-tuned configuration. Comparisons between an actively detuned coil and a purely geometrically decoupled architecture were used to demonstrate the capabilities of the design. In particular cases, the simplicity and adaptability of the interleaved nine-leg design for multinuclear nuclear magnetic resonance (NMR) offer a straightforward alternative to the often complex and lossy designs currently available for multinuclear birdcages and volume coils.

Citation format

BUSHER, Joseph, et al. Odd-leg birdcages for geometric decoupling in multinuclear imaging and spectroscopy. CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2023, 2023.