PhysicsEngineering

Qingqian Guo, Chang-yu Ma, Xin Zhang, Yajie Xu, Meisheng Fan, P. Yu, Tao Hu, Yan Chang, Xiao-dong Yang

2020.10.22CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING

DOI: 10.1155/2020/8882329

tlooto Summary

SQUID-based MRI system developed with ultra-low field strength and backprojection method for high-resolution imaging at microtesla range.

Abstract

Ultra-low field magnetic resonance imaging (ULF MRI) is an effective imaging technique that applies the ultrasensitive detector of superconducting quantum interference device (SQUID) sensor to detect the MR signal at a microtesla field range. In this work, we designed and developed a SQUID-based ULF MRI system with a frequency-adjustable measurement field, the performance of which was characterized via water phantoms. In order to enhance the MR signals, a 500 mT Halbach magnet was used to prepolarize the magnetization of the sample prior to excitation. The signal-to-noise-ratio (SNR) of the spin-echo- (SE-) based pulse sequence can reach up to 70 in a single scan. The images were then reconstructed successfully by using the maximum likelihood expectation maximization (MLEM) algorithm based on the backprojection imaging method. It was demonstrated that an in-plane resolution of 1.8 × 1.8 mm2 can be achieved which indicated the feasibility of SQUID-based MRI at the ULF.

Citation format

GUO, Qingqian, et al. SQUID-Based magnetic resonance imaging at ultra-low field using the backprojection method. CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2020, 2020: 1–11.